OPERATION
63
dBm/dB/W Measurement
dBm/dB Calculation
Overview
Using the ACV or DCV measurement results,
the DMM calculates the dB or dBm value based
on a reference resistance value in the following
way:
dBm= 10 x log
10
(1000 x Vreading
2
/ Rref)
dB= dBm dBmref
W= Vreading
2
/Rref
Where:
Vreading= Input Voltage, ACV or DCV;
Rref= Reference resistance simulating an output
load;
dBmref= Reference dBm value
Measuring dBm/W
Procedure
1. Select ACV or DCV measurement. See page 32.
2. To measure dBm, press SHIFT
MENU(dBm)
The primary display will show the dBm
measurement while the secondary display
shows the reference resistance.