EasyManua.ls Logo

GW Instek GDM-8351 - Normal Mode

GW Instek GDM-8351
159 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GDM-8351 User Manual
90
Wiring Diagram
Pins 1 ~ 4
Pins 1 ~ 4 are open-collector outputs, with a
max input of 30mA. All outputs are active low.
Pin 1-4
Vo
Imax = 30mA
Digital I/O
Terminal
Ext Vcc
Wiring Diagram
Trig In (Pin 5)
Pin 5
Pin 6
Trigger input
Digital I/O
Terminal
Normal Mode
Overview
The Normal Mode outputs the pass/fail results
of the Compare function. Each signal is an
active low signal. In addition an active low
pulse of approximately s is output to indicate
the end of compare measurement (EOM).
When the input signal exceeds the high
threshold or the low threshold, the High Fail or
Low Fail pin is pulled low. When the signal
stays within the threshold levels, the Pass pin is
pulled low.

Table of Contents

Other manuals for GW Instek GDM-8351

Related product manuals