REMOTE CONTROL
295
[SENSe:]FRESistance:TRIGger:DELay?
Returns the trigger delay time in seconds of 4-wire resistance measurement.
Return parameter: <NRf>
[SENSe:]FRESistance:ZERO:AUTO
Sets the auto zero mode to on, off or once of 4-wire resistance
measurement.
Parameter: 0 | 1 | ON | OFF | ONCE
Example: SENS:FRES:ZERO:AUTO ON
Sets the auto zero to on.
[SENSe:]FRESistance:ZERO:AUTO?
Returns the auto zero mode of 4-wire resistance measurement.
Return Parameter: 0 | 1, 1=ON, 0=OFF
SENSe FREQency Commands
[SENSe:]FREQuency:APERture
Sets the aperture time (gate time) for the frequency function (0.01s, 0.1s,
1s).
Parameter: <NRf> (0.01 | 0.1 | 1)
Example: SENS:FREQ:APER 0.01
Sets the gate time to 0.01 seconds.
[SENSe:]FREQuency:APERture?
Returns aperture time (gate time) for the frequency function.
Return parameter: <NRf>
[SENSe:]FREQuency:CURRent:RANGe
Sets the frequency measurement range.
Parameter: Range(<NRf> | MIN | MAX | DEF)
Example: SENS:FREQ:CURR:RANG MIN
Sets the frequency to the minimum range.
[SENSe:]FREQuency:CURRent:RANGe?
Returns the frequency measurement range.
[SENSe:]FREQuency:CURRent:RANGe:AUTO
Sets the Frequency Auto-range settings on, off or once.
Parameter: 0 | 1 | ON | OFF | ONCE
Example: SENS:FREQ:CURR:RANG:AUTO ON
Turns the Auto-range on for the frequency function.
[SENSe:]FREQuency:CURRent:RANGe:AUTO?
Returns the frequency Auto-range setting.
Return parameter: 0 | 1, 0=OFF, 1=ON