REMOTE CONTROL
297
[SENSe:]FREQuency:TIMeout:AUTO?
Returns the assigned timeout time used for the frequency function.
Return parameter: 0 | 1, 0:timeout time = 1 second, 1:timeout time is
different in according with ac filter bandwith (gate time).
[SENSe:]FREQuency:TRIGger:DELay
Sets the trigger delay time that minimum step is microseconds of
Frequency measurement.
Parameter: <NRf> (0 ~ 3600 s) | MIN | MAX | DEF
Example: SENS:FREQ:TRIG:DEL 0.5
Sets the trigger delay time to 0.5s of Frequency measurement.
[SENSe:]FREQuency:TRIGger:DELay?
Returns the trigger delay time in seconds of Frequency measurement.
Return parameter: <NRf>
[SENSe:]FREQuency:VOLTage:RANGe
Sets the frequency measurement range.
Parameter: Range(<NRf> | MIN | MAX | DEF)
Example: SENS:FREQ:VOLT:RANG MIN
Sets the frequency to the minimum range.
[SENSe:]FREQuency:VOLTage:RANGe?
Returns the frequency measurement range.
[SENSe:]FREQuency:VOLTage:RANGe:AUTO
Sets the Frequency Auto-range settings on, off or once.
Parameter: 0 | 1 | ON | OFF | ONCE
Example: SENS:FREQ:VOLT:RANG:AUTO ON
Turns the Auto-range on for the frequency measurement.
[SENSe:]FREQuency:VOLTage:RANGe:AUTO?
Returns the Frequency Auto-range setting.
Return parameter: 0 | 1, 0=OFF, 1=ON
[SENSe:]PERiod:APERture
Sets the aperture time (gate time) for the period function(0.01s, 0.1s, 1s).
Parameter: <NRf> (0.01 | 0.1 | 1)
Example: SENS:PER:APER 0.1
Sets the gate time to 0.1 seconds for the period function.
[SENSe:]PERiod:APERture?
Returns the aperture time (gate time) for the period function.
Return parameter: <NRf>