REMOTE CONTROL
299
[SENSe:]PERiod:NULL:VALue:AUTO
Sets the relative value auto on/off for Period measurement.
Parameter: 0 | 1 | ON | OFF
Example: SENS:PER:NULL:STAT ON
SENS:PER:NULL:VAL:AUTO ON
The unit automatically sets the 1st count of measurement as null value.
[SENSe:]PERiod:NULL:VALue:AUTO?
Returns the null value auto state of Period measurement.
[SENSe:]PERiod:TIMeout:AUTO
Assigns timeout time at the period measurement.
Parameter: 0 | 1 | ON | OFF
Example: SENS:PER:TIM:AUTO ON
Sets the timeout time in according with ac filter bandwith (gete time).
[SENSe:]PERiod:TIMeout:AUTO?
Returns the assigned timeout time used for the period function.
Return parameter: 0 | 1, 0:timeout time = 1 second, 1:timeout time is
different in according with ac filter bandwith (gate time).
[SENSe:]PERiod:TRIGger:DELay
Sets the trigger delay time that minimum step is microseconds of Period
measurement.
Parameter: <NRf> (0 ~ 3600 s) | MIN | MAX | DEF
Example: SENS:PER:TRIG:DEL 0.05
Sets the trigger delay time to 50ms of Period measurement.
[SENSe:]PERiod:TRIGger:DELay?
Returns the trigger delay time in seconds of Period measurement.
Return parameter: <NRf>
[SENSe:]PERiod:VOLTage:RANGe
Sets the period measurement range.
Parameter: Range(<NRf> | MIN | MAX | DEF)
Example: SENS:PER:VOLT:RANG DEF
Sets the period to the default range.
[SENSe:]PERiod:VOLTage:RANGe?
Returns the period measurement range.
[SENSe:]PERiod:VOLTage:RANGe:AUTO
Sets the Period Auto-range settings on, off or once.
Parameter: 0 | 1 | ON | OFF | ONCE
Example: SENS:PER:VOLT:RANG:AUTO OFF
Turns the Auto-range setting off for period measurements.