GDM-906X Series User Manual
300
[SENSe:]PERiod:VOLTage:RANGe:AUTO?
Returns the Period Auto-range setting.
Return parameter: 0 | 1, 0=OFF, 1=ON
SENSe TEMPerature Commands
[SENSe:]TEMPerature:NPLCycles
Sets the integration time for Temperature measurement in PLCs (power line
cycles). Where one PLC is equal to 16.6 milliseconds. For any <NRf>
parameter, the DMM will automatically set the PLC to the closest
acceptable PLC value (1 | 3 | 12).
Parameter: NPLCycles(<NRf> | MIN | MAX | DEF)
Example: SENS:TEMP:NPLC DEF
Sets the integration time to the 12 PLCs for Temperature measurement.
[SENSe:]TEMPerature:NPLCycles?
Returns the integration time for Temperature measurement in PLCs (power
line cycles). Where one PLC is equal to 16.6 milliseconds.
Return parameter: 1 | 3 | 12
[SENSe:]TEMPerature:NULL[:STATe]
Sets the relative function on/off for Temperature measurement.
Parameter: 0 | 1 | ON | OFF
Example: SENS:TEMP:NULL:STAT ON
Turns the relative function on for Period measurement.
[SENSe:]TEMPerature:NULL[:STATe]?
Returns the relative function state of Temperature measurement.
Return parameter: 0 | 1, 0=OFF, 1=ON
[SENSe:]TEMPerature:NULL:VALue
Sets the relative value for Temperature measurement.
Parameter: <NRf> (-1.0e15~1.0e15) | MIN | MAX | DEF
Example: SENS:FREQ:NULL:VAL 5
Sets the relative value to 5˚C for Temperature measurement.
[SENSe:]TEMPerature:NULL:VALue?
Returns the current relative value of Temperature measurement.
[SENSe:]TEMPerature:NULL:VALue:AUTO
Sets the relative value auto on/off for Temperature measurement.
Parameter: 0 | 1 | ON | OFF
Example: SENS:TEMP:NULL:STAT ON
SENS:TEMP:NULL:VAL:AUTO ON
The unit automatically sets the 1st count of measurement as null value.