REMOTE CONTROL
301
[SENSe:]TEMPerature:NULL:VALue:AUTO?
Returns the null value auto state of Temperature measurement.
[SENSe:]TEMPerature:RESolution
Sets the Temperature measurement resolution. The resolution depends on
the rate and range settings.
Parameter: Resolution(<NRf> | MIN | MAX | DEF)
Example: SENS:TEMP:RES MAX
Sets the Temperature resolution to the maximum.
[SENSe:]TEMPerature:RESolution?
Returns the temperature measurement resolution.
[SENSe:]TEMPerature:TRANsducer:TYPE
Sets the temperature probe type.
Parameter: [None] | TC | RTD | FRTD | THER | FTH
Example: SENS:TEMP:TRAN:TYPE RTD
Sets the temperature probe type to RTD.
[SENSe:]TEMPerature:TRANsducer:TYPE?
Returns the temperature probe type.
Return parameter: TC, RTD, FRTD, THER, FTH
[SENSe:]TEMPerature:TRIGger:DELay
Sets the trigger delay time that minimum step is microseconds of
Temperature measurement.
Parameter: <NRf>(0 ~ 3600 s) | MIN | MAX | DEF
Example: SENS:TEMP:TRIG:DEL 0.001
Sets the trigger delay time to 1ms of Temperature measurement.
[SENSe:]TEMPerature:TRIGger:DELay?
Returns the trigger delay time in seconds of Temperature measurement.
Return parameter: <NRf>
[SENSe:]TEMPerature:ZERO:AUTO
Sets the auto zero mode to on, off or once of Temparature measurement.
Parameter: 0 | 1 | ON | OFF | ONCE
Example: SENS:TEMP:ZERO:AUTO OFF
Sets the auto zero to off.
[SENSe:]TEMPerature:ZERO:AUTO?
Returns the auto zero mode of Temparature measurement.
Return Parameter: 0 | 1, 1=ON, 0=OFF