BASIC MEASUREMENT
43
10kΩ
10mΩ
11.99999
100kΩ
100mΩ
119.9999
1MΩ
1Ω
1.199999
10MΩ
10Ω
11.99999
100MΩ
100Ω
119.9999
Note
For more details, see the specifications on page 356.
General Resistance Setting
F2 (Speed) key to
select the rate
Press the F1 ~ F5 key to select the desired rate
Press the F6 (More 1/2) key for next page with
more options as the figure shown below.
F3 (AutoZero) key
to enable Auto Zero
Background
Autozero provides the most
accurate measurements, but
requires additional time to
perform the zero measurement.
With autozero enabled (On), the
GDM-9060/9061 internally
measures the offset following
each measurement. It then
subtracts that measurement from
the preceding reading. This
prevents offset voltages present
on the GDM-9060/9061’s input
circuitry from affecting
measurement accuracy. With
autozero disabled (Off), the
GDM-9060/9061 measures the
offset once and subtracts the
offset from all subsequent
measurements.