BASIC MEASUREMENT
61
RTD 2W/4W Setting
Background
The GDM-9060/9061 supports 2 or 4 wire RTD. It is
important to specify the type of temperature sensor used.
Parameter
RTD type
Range
Resolution
All (based on PT100)
-200~630°C
0.001°C
Procedure
1. Press the F1 (Probe) key to enter the Temperature
Probe menu followed by clicking either the F2 (RTD 2W)
or F3 (RTD 4W) key to activate RTD
2W/4W mode. See the figure shown below.
2. Press the F5 (Type) key to enter the Sensor Type menu
as the figure shown below. Click the F1 F5 key to select a
desired sensor type per your requirement.
3. The display shows the latest setting. See the example of the
figure below where RTD 2W: PT100 is currently activated by
user.
Probe: RTD 2W
Type: PT100
Temperature
Measurement
R0 Value
w ww. . com