SAFETY INSTRUCTIONS
7
If the equipment is used in a manner not specified by the
manufacturer, the protection provided by the equipment
may be impaired. Do NOT replace the detachable
MAINS supply cords by inadequately RATED cords.
Suitable supply cord set for use with the equipment:
Mains plug: Shall be national approval
Mains connector: C13 type
Cable:
1. Length of power supply cord: less than 3 m
2. Cross-section of conductors: at least 0.75 mm2
3. Cord type shall meet the requirements of IEC
60227 or IEC 60245 (e.g.: H05VV-F, H05RN-F)
Due to the fact that the Front/Rear Input Switch on
the front panel is not proposed as an active
multiplexer, do Not change the input switch when
signals are present on either rear or front set of
terminals. Intrument damage and risk of electric shock
may occur if switching the input switch when high
voltage or current is present.
Fuse type: T0.25A 100/120 VAC
T0.125A 220/240 VAC
Make sure the correct type of fuse is installed before
power up.
To avoid risk of fire, replace the fuse only with the
specified type and rating.
Disconnect the power cord before fuse replacement.
Make sure the cause of a fuse blowout is fixed before
fuse replacement.
Disconnect the power cord before cleaning.
Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid into the
GDM-9060/9061.
Do not use chemicals or cleaners containing harsh
material such as benzene, toluene, xylene, and acetone.