GPT-9000/9000A Series User Manual
142
Example
(GPT-99XX/
99XXA)
MANU:IR:RHIS 0.010
Sets the IR HI SET resistance to 10 M.
MANU:IR:RLOSet
Set
Query
Description
Sets or returns the IR LO SET resistance value in
M (GPT-98XX) or GΩ. The LO SET value must be
less than the HI SET value. The test must first be
in IR mode before this command can be used.
Syntax
Query Syntax
MANU:IR:RLOSet<NR1>
MANU:IR:RLOSet?
Parameter/
Return parameter
<NR1>
GPT-98XX only:
1 ~ 9999 (unit = M)
GPT-99XX/GPT-99XXA only:
Format A: 0.001 ~ 50.00 (unit = G)
Format B: 0.001G ~ 50.00G
Format C: 1M ~ 50000M
Example
(GPT-98XX)
MANU:IR:RLOS 10
Sets the IR LO SET resistance to 10M.
Example
(GPT-99XX/
99XXA)
MANU:IR:RLOS 0.010
Sets the IR LO SET resistance to 10M.
MANU:IR:TTIMe
Set
Query
Description
Sets or returns the IR test time in seconds. The test
must first be in IR mode before this command can
be used.
Syntax
Query Syntax
MANU:IR:TTIMe <NR2>
MANU:IR:TTIMe?
Parameter/
Return parameter
<NR2>
1.0 ~ 999.9 seconds
Example
MANU:IR:TTIM 1
Sets the IR test time to 1 second.
w ww . . co m
information@itm.com1.800.561.8187
All manuals and user guides at all-guides.com

Table of Contents

Other manuals for GW Instek GPT-9000 series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GPT-9000 series and is the answer not in the manual?

GW Instek GPT-9000 series Specifications

General IconGeneral
BrandGW Instek
ModelGPT-9000 series
CategoryTest Equipment
LanguageEnglish

Related product manuals