LCR-1000 Series
20
Battery installation and replacement
The instrument has a built-in rechargeable lithium battery, which
has been installed in the battery compartment of the instrument at
the factory. If replacing the battery, follow the steps below
1. Squeeze the axis of the instrument bracket
inward, and remove the bracket.
2. Use a screwdriver to loosen the 3 screws on
the battery cover and remove the battery
cover.
3. Remove the plug on the old battery, plug in
the plug of the new battery, pay attention to
the direction of the plug.
4. Put the new battery into the battery
compartment, cover the battery cover, and
tighten the screws.
5. Install the instrument bracket.
Figure
Installing and
Replacing the
Battery
Warning
Please use the original lithium battery (LCR-305)
to ensure the safety of the instrument. Using non-
original batteries may cause danger to the user or
damage the instrument.