OPEN/SHORT
33
Open test
Open correction capability cancels errors due to the stray
admittance(G,B) in the test fixture.
Figure
Stray admittance
Function
ON/OFF
Enable or Disable the open circuit
correction function.
MEAS OPEN
Open correction for LCR only.
DCR OPEN
Open correction for DCR only.
Steps
1. Press the [OPEN SHORT] key to enter the
[OPEN SHORT] page.
2. Use the arrow keys to select the [OPEN
TEST] item.
3. Select [MEAS OPEN] or [DCR OPEN] to
start open correction.
LCD display prompt:
Open-circuit the
test terminals
At this time, please keep the test terminal
or test fixture open and do not contact with
any objects.