OPEN/SHORT
33
Open test
Open correction capability cancels errors due to the stray
admittance(G,B) in the test fixture.
Enable or Disable the open circuit
correction function.
Open correction for LCR only.
Open correction for DCR only.
1. Press the [OPEN SHORT] key to enter the
[OPEN SHORT] page.
2. Use the arrow keys to select the [OPEN
TEST] item.
3. Select [MEAS OPEN] or [DCR OPEN] to
start open correction.
LCD display prompt:
Open-circuit the
test terminals
At this time, please keep the test terminal
or test fixture open and do not contact with
any objects.