OPEN/SHORT
35
Short test
Short correction compensates for any residual impedances (R,X),
such as the impedance of the cables and the DUT connection
points.
Figure
Residual
impedances
ON/OFF
Enable or Disable the short
circuit correction function.
MEAS SHORT
Short correction for LCR only.
DCR SHORT
Short correction for DCR only.
Steps
1. Press the [OPEN SHORT] key to enter
the [OPEN SHORT] page.
2. Use the arrow keys to select the
[SHORT TEST] item.
3. Select [MEAS SHORT] or [DCR SHORT] to
start short correction.
LCD display prompt:
Short-circuit the test
terminals
Please insert the short bar into the test
terminal or short the test fixtures together.