LCR-1000 Series
40
Touch screen
calibration steps
1. Touch screen calibration requires a touch
pen, which can be calibrated with a mobile
phone/tablet touch pen.
It is not recommended to use fingers to
calibrate the touch screen, because the
finger contact point is large and inaccurate.
Press the CALIBRATE function key to enter
the calibration page. Use the touch pen to
touch the screen, and the calibration starts.
2. Calibrate the first point: touch the cross in
the upper left orner.