57
Command List
Summary of all USB commands
Command (sHeader)
Parameter (sPara)
Description
DISP:PAGE
MEAS|SETUP|SYSTem|CSET
|SINF
Switch display page
DISP:PAGE?
Query the name of the
currently displayed page
DISP:LINE
String(maximum 28
characters)
Display the string in the
bottom prompt bar
FUNC
C-D|C-Q|C-R|L-D| L-Q|L-R|L-
Rdc| R-Q|R-X|R-dc|Rdc|
Z-D|Z-Q|Z-thr|Z-thd
Set the measurement
function
FUNC?
Query main parameter
options
FUNC:EQU
SERIAL|PARALLEL|AUTO
Set the equivalent mode
option
FUNC:EQU?
Query equivalent mode
options
FUNC:LCR:RANG
0~7
Set LCR range
FUNC:LCR:RANG?
Query the LCR range
number
FUNC:DLCR:RANG
0~7
Set DCR range
FUNC:DCR:RANG?
Query the DCR range
number
FUNC:RANG:AUTO
ON|OFF|1|0
Set the range automatically
FUNC:RANG:AUTO?
Query the automatic state
of the range
FREQ
50|100|120|1k|2k|10k|50k|10
0k
set the frequency value
FREQ?
query frequency value
APER
SLOW | FAST
Set test speed
APER?
Query test speed
FETC?
Query test results
COMP
ON | OFF | 1 | 0
Turn on/off the comparator
COMP:BEEP
OFF | PASS | FAIL
Turns the beep on/off
COMP:NOM
<float number>
Enter nominal value
COMP:NOM?
Query the nominal value
COMP:TOL
<float number>
Enter a percentage value
COMP:TOL?
Query percentage value

Related product manuals