LCR-1000 Series
62
FREQ <50|100|120|1k|2k|10k|50k|100k>
FREQ?
Function It sets or queries
the test frequency.
Description
Set parameter <50|100|120|1k|2k|10k|50k|100k>
Set syntax FREQ
Query syntax FREQ?
Return data <50|100|120|1000|2000|10000|50000| 100000>
Note
The frequency value can accept any numerical
format, but the data must be the frequency
value of the instrument, and the incorrect
frequency value will use the closest correct
frequency value.
The frequency values for each model are as
follows:
LCR-1010: 50|100|120|1k|2k|10k
LCR-1100:
50|100|120|1k|2k|10k|50k|100k
APER <SLOW|FAST>
APER?
Function It sets or queries
the test speed.
Description
Set parameter <SLOW|FAST>
Set syntax APER
Query syntax APER?
Return data <SLOW|FAST>
FETC?
Function It queries
the test results.
Description
Query syntax FETC?
Return data <primary parameter result, secondary parameter