COMMAND OVERVIEW
63
result>
Note
The instrument returns the test results of the
primary and secondary parameters NR3 type.
If the secondary parameter is closed, it will
return data +0.000000e+00
Example: +7.929158e-15,+0.000000e+00
COMP <ON | OFF | 0 | 1>
COMP?
Function It sets or queries
the status of the comparator.
Description
Set parameter <ON | OFF | 0 | 1>
Set syntax COMP
Query syntax COMP?
Return data <on|off>
COMP:BEEP <OFF | PASS | FAIL>
COMP:BEEP?
Function It sets or queries
the status of the beep.
Description
Set parameter <OFF | PASS | FAIL>
Set syntax COMP:BEEP
Query syntax COMP:BEEP?
Return data <OFF | PASS | FAIL>
COMP:NOM <NR1|NR2|NR3|NR4>
COMP:NOM?
Function It inputs and queries the data of nominal value.
Description
Set parameter <NR1|NR2|NR3|NR4> any form of data
Set syntax COMP:NOM
Query syntax COMP:NOM?
Return data <NR3> Example: 1.000000e-09