LCR-1000 Series
68
SPECIFICATIONS
Specifications
If the following conditions are met during the measurement, the
measurement result can reach the basic accuracy.
Warranted performance. All specification apply at 23 ± 5 °C, unless
otherwise stated, and 30 minutes after the instrument has been
turned on.
Calibration: Yearly
Reset adjustment: Perform correction before testing
the test port of the instrument itself
Test cable length: 0m
Open and short calibration has been performed
When using the external power, the accuracy is unspecified.
General Characteristics
Specification Conditions:
Temperature: 18˚C~28˚C
Humidity: 70%RH(non-condensing)
Operating Environment
Temperature Range: 0~40˚C
Relative Humidity: 80%RH(non-condensing)
Storage Conditions
Temperature Range: -10~70˚C
Relative Humidity: 90%RH(non-condensing)
Power
Power: Rechargeable lithium battery(8.4V)