Accuracy
134
13.1.3 Accuracy for Q
Q’s accuracy is defined below:
Q
e
=
e
D
x
Q1
e
D
2
x
Q
(when Q
x
×D
e
<1)
Where:
Q
x
is the measured Q value.
D
e
is D’s accuracy.
13.1.4 Accuracy for
s accuracy is defined below:
100
180 e
e
A
[deg]
13.1.5 Accuracy for Rp
When D
x
(the measured D value)≤0.1, R
p
’s accuracy is
defined as:
R
p
=
e
D
x
D
e
D
px
R
[Ω]
Where:
R
px
is the measured R
p
value [Ω].