Measure Key
42
4.3 [LIST SETUP] Page
Press the [Measure] key and press the [LIST SETUP]
soft-key to open the [LIST SETUP] page. The List
Measure feature can perform automatic sweep
measurements by sweeping the frequency or signal level
through a maximum 10 list points.
Before using the List Measure feature, you have to
configure the list setup.
In the [LIST SETUP] page, you can configure each of
the following list measurement controls with the cursor
placed in the corresponding field.
Function mode [FUNC]
Test mode [MODE]
Parameter selection [FREQ[Hz], VOLT[V],
CURR[A]]
Limit parameter selection [LMT]
Lower and upper limits [LOWER] [UPPER]
Figure 4-5 [LIST SETUP] Page
4.3.1 Measurement Function [FUNC]
Select the primary and the secondary measurement here.