LCR-8000 Series User Manual
104
Setting OUTPUT IMPEDANCE
The output impedance can be set to 25Ω or 100Ω.
The varied signal source output impedance will lead to the varied
current or the difference of measuring value. If selecting <25Ω>,
then the voltage range is 10mV~1Vrms and the current range is
400µA~40mArms.
If need to compare test results with Keysight select 100Ω.
1. Press the Measure button to enter
[MEAS DISPLAY] page, and press the
LIST SET function button on the right
side of the LCD to enter the [LIST SET]
page, and then press the Setup button
to enter [LIST MODE SETUP] page.
2. Use arrow keys to move cursor and
select OUTPUT IMPEDANCE item on
the [LIST MODE SETUP] page.
3. Use option key on the right of the LCD screen to
select an impedance item.
Set output impedance 100Ω