CORRECTION (OPEN/SHORT)
129
Setting spot load
This item does not necessarily need to be implemented. If the
OPEN/SHOR/LOAD compensation cannot make the
measurement correct, you can perform SPOT LOAD compensation
to let the instrument know the correct value through learning.
Use the part with the known value as the standard value which will
be recognized by the instrument during the test, the instrument
will then shows this value in the test status.
Steps
1. Press the Measure button to enter
[MEAS DISPLAY] page, and then
press the CORR. function button on
the right side of the LCD to enter the
[CORRECTION] page.
2. Use arrow keys to move cursor and
select SPOT LOAD item on the
[CORRECTION] page.
3. Use option key on the right of the LCD screen
to select OFF/ON item.
4. When set to ON, the setting items of the SPOT
No. below will be turned enable.