LCR-8000 Series User Manual
132
Setting fixture compensation
If use a dedicated test fixture, select the corresponding option to do
OPEN/SHORT/LOAD correction, which can increase the
measurement accuracy.
Steps
1. Press the Measure button to enter
[MEAS DISPLAY] page, and then
press the CORR. function button on
the right side of the LCD to enter the
[CORRECTION] page.
2. Use arrow keys to move cursor and
select FIXTURE COMPENSATION
item on the [CORRECTION] page.
3. Use option key on the right of the LCD screen
to select FIXTURE item.
Note
Only when it is used for test fixtures above 3MHz,
it is necessary to set the fixture compensation.