LCR-8000 Series User Manual
186
:MEASure:DELay <delay time NR3/disc>
:MEASure:DELay?
Function: It sets or queries the AC/DC DELAY.
Description:
Set parameter The value of delay time
is
{0.0~5|MAXimum|MINimum}.
Set syntax
:MEASure:DELay 0.5
:MEASure:DELay 500M
:MEASure:DELay 500MS
:MEASure:DELay 5E-3
:MEASure:DELay MAXimum
:MEASure:DELayMINimum
Query syntax
:MEASure:DELay?
Return
data 0.500 (Format is in <NR2>)
:MEASure:TRIGger:MODE {REPeat|0|SINGle|1|}
Function: It sets or queries the TRIGGER MODE.
Description:
Set parameter REPeat,0
(repeat),
SINGle,1 (single)
Set syntax
:MEASure:TRIGger:MODE SINGle
Query syntax
:MEASure:TRIGger:MODE?
Return data REP | SING (Format is in <disc>)
:MEASure:RANGe:DC {1|2|3|4|AUTO|HOLD}
:MEASure:RANGe:DC?
Function: It sets or queries the DC RANGE.
Description:
Set parameter {1|2|3|4|
AUTO
|
HOLD
}
Set syntax
:MEASure:RANGe:DC 2
:MEASure:RANGe:DC AUTO
Query syntax :
MEASure:RANGe:DC
Return Data 2 (Format is in <NR1>)