LCR-8000 Series User Manual
196
:LIST:FREQuency 1KHZ
:LIST:FREQuency 1E3
:LIST:FREQuency MAXimum
:LIST:FREQuencyMINimum
Query syntax
:LIST:FREQuency?
Return data 1.000000E+03(Format is in <NR3)
:LIST:VOLTage<voltage NR3/disc >
:LIST:VOLTage?
Function: It sets or queries the measurement voltage of list mode.
Description:
Set parameter When RO(output impedance) is 100Ω, the setup
range of AC voltage is {0.01~2|MAXimum|MINimum}. When
RO(output impedance) is setting 25Ω, the setup range of AC
voltage is {0.01~1|MAXimum|MINimum}. The range of DC voltage
is {0.01~1|MAXimum|MINimum}.
Set syntax
:LIST:VOLTage 1
:LIST:VOLTage 1M
:LIST:VOLTage 1MV
:LIST:VOLTage 1E-3
:LIST:VOLTage MAXimum
:LIST:VOLTageMINimum
Query syntax
:LIST:VOLTage?
Return data 1.000000E+00 (Format is in <NR3>)
When the Level setup is current mode, the data returns 9.9E37.
:LIST:CURRent <current NR3/disc >
:LIST:CURRent?
Function: It sets or queries the measurement current of list mode.
Description:
Set parameter when RO(output impedance) is 100Ω, the setup
range of AC current is {0.0002~0.02|MAXimum|MINimum}. When
RO(output impedance) is setting 25Ω, the setup range of AC