GW Instek LCR Series - Sweep:traca:position {-10 ~ 16|Maximun|Minimum; Sweep:traca:position; Sweep:traca:division <Value Nr3;Disc; Sweep:traca:division

GW Instek LCR Series
250 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LCR-8000 Series User Manual
214
:SWEep:TRACA:POSition {-10 ~ 16|MAXimun|MINimum}
:SWEep:TRACA:POSition?
Function It sets or queries the position of TRACE A.
Description
Set parameter
{-10 ~ 16|MAXimun|MINimum}
Set syntax
:SWEep:TRACA:POSition -3
Query syntax
:SWEep:TRACA:POSition?
Return
data -3 (Format is in <NR1>)
:SWEep:TRACA:DIVision <value NR3/disc>
:SWEep:TRACA:DIVision?
Function It sets or queries the division of TRACE A.
Description
Set parameter The numeric value can only set 1, 2, 5 and 10(when
Y-axis setup is linear mode).
Set syntax
:SWEep:TRACA:DIVision 2k
Query syntax
:SWEep:TRACA:DIVision?
Return data 2.000000E+03 (Format is in <NR3>)
:SWEep:TRACA:DECade {1~12}
:SWEep:TRACA:DECade?
Function It sets or queries the decade of TRACE A.
Description
Set parameter The value of DECADE is {1~12} (when Y-axis setup
is LOG mode).
Set syntax
:SWEep:TRACA:DECade 5
Query syntax
:SWEep:TRACA:DECade?
Return
data 5 (Format is in <NR3>).