Setting measurement level/ALC/RO
LCR-8200 Series test signal voltage/current level can be set as the
effective value (RMS value) of a sine wave of the test frequency
from the unit’s internal oscillator. The voltage range is 10mV-
2Vrms and current range is 100uA-20mArms. 2Vrms can only be
used at <= 1MHz.
The ALC (automatic level control) feature adjusts the voltage across
the DUT or the current through the DUT to match the
voltage/current level setting. Using this feature, you can try to
ensure a constant signal level (voltage or current) is applied to the
DUT.
In situations when the actual measuring Vac or Iac goes beyond the
extent that the ALC can regulate, a warning message, “ALC unable
to regulate”, will be shown at the bottom of the screen.
An asterisk will be shown beside the LEVEL V or A unit when the
ALC is turned on.
The RO (output impedance) can be set to 2 or 100Ω.
The varied signal source output impedance will lead to the varied
current or the difference of measuring value. If selecting <25Ω>,
then the voltage range is 10mV~1Vrms and the current range is
400uA~40mArms. If it needs to compare test results with Keysight
select 100Ω.