APPENDIX
257
DYNAMIC MODE
CONSTANT CURRENT MODE
T1&T2 0.025mS ~ 10mS / Res : 1uS
10mS ~ 30S / Res : 1mS
Accuracy 1uS / 1mS + 100ppm
Slew Rate 0.8 ~ 200mA/uS 0.64 ~ 160mA/uS
6.4 ~
1600mA/uS
Slew Rate
Resolution
0.8mA/uS 0.64mA/uS 6.4mA/uS
Slew Rate setting
Accuracy
±(10% + 15us) ±(10% + 15us) ±(10% + 15us)
Current 0~5A 0~4A 0~40A
Current
Resolution
1mA 1mA 10mA
Current Accuracy ±0.4% F.S.
MEASUREMENT
VOLTAGE READBACK
Range 0~16V 0~80V 0~16V 0~80V
Resolution 0.5mV 2.5mV 0.5mV 2.5mV
Accuracy ±(0.05%set + 0.05% F.S.)
CURRENT READBACK
Range 0~5A 0~4A 0~40A
Resolution 0.15625mA 0.125mA 1.25mA
Accuracy ±(0.1%set + 0.1% F.S.)
PROTECTIVE
Over Power Protection
Range 0.9~30.6W 1.25~255W
Resolution 0.15W 1.25W
Accuracy ±(2%set +
0.25%F.S)
±(2%set + 0.25%F.S)
Over Current Protection
Range 0.0625~5.1A 0.5~40.8A
Resolution 0.0125A 0.1A
Accuracy ±(2%set +
0.25%F.S)
±(2%set + 0.25%F.S)
Over Voltage Protection
Range 1~81.6V 1~81.6V
Resolution 0.2V 0.2V
Accuracy ±(2%set +
0.25%F.S)
±(2%set + 0.25%F.S)