PEL-3000(H) User Manual
148
Last_C
Step_T
Delay
Trig_V
4. Press the Save[F3] to save the selected test
setup.
Start OCP
5. Press OCP[F1] to turn the OCP function on if it
is off.
6. The OCP function can be started by turning the
load on by pressing
Shift Load
On/
Off
.
The test current will increase from the Start C
value to the End C value in steps according to
the Step C value, until the test has finished.
The test will start running when the power
supply voltage is greater than the Trig V
voltage.
Example: OCP
Function running
OCPRS232
01/Oct/2017
OCP Test
7.498
V
0.1531
A
1.15
W
Set Curr: Fetch Volt:
2.000
1.500
1.000
6.91
6.92
6.91
Measured voltage for
last three steps
Measured voltage,
current and power
Running.
Set current for the
last three steps
(descending order)