6. Set the power level parameters using the scroll
wheel and number pad.
For Static mode, set CP A Value and/or CP B
Value.
For Dynamic mode, set Level1 and Level2.
The maximum and minimum power levels
depend on the selected current range.
For static mode, the parameter that is set last
becomes the “active” setting. This will be
shown in the Operation Status Panel.
7. To add CV mode to CP mode (CP+CV), see
page 60.
8. Set the remaining basic configuration settings
such as the slew rate, and timer settings. See
page 65 for details.
Basic CP mode configuration is complete. See page 65
for more configuration options.
The current range and voltage range applies to all the
operating modes.