PEL-3000(H) User Manual
156
Power Source
Config Error
Exit
OPPUSB
01/Oct/2018
OPP Test
4.90
V
0.000
A
0.00
W
Fetch Watt: Fetch Volt:
0.00 4.91
Config Error
Config Error indicates that the power supply
voltage is less than the Trig V voltage setting after
the test has started. This can indicate that the
power supply output is not on or that the power
supply output or Trig V is incorrectly configured.
In addition to the OPP settings as described above,
the Trig voltage settings must also be set according to
the output characteristics of the DUT.