OPERATION
69
Slew Rate
Description
The current slew rate can be set for CC and CR
mode. The slew rate setting is used to limit the
change in current when switching.
For static mode, only a single slew rate can be set.
Operation
1. Make sure the load is off.
2. Press
Main
.
3. Set the slew rate(s) using the scroll wheel and
number pad.
For static mode, only a single slew rate can be
set.
For dynamic mode, set both the rising and
falling slew rates.
Take the timer settings into consideration when
setting the slew rates.
Display
Mode
CC
I Range
H 8.75A
V Range
H 800V
Function
Dynamic
Configure
0.00
V
0.0000
A
0.00
W
Timer1 0.025
SlewRate 140.00
SlewRate 140.00
LOADRS232
01/Oct/2018
ms
mA/us
mA/us
Slewrate settingsSlewrate settings