Panel Operation
47
Output Current Setting
Operation
1. Press the Vset/ Iset key
VSET/ ISET
repeatedly to move the underline
to the Current side.
OCP
SET 2.58V 2
.01A
2. Press the F/C key
F/C
to move the
underline before or after the
decimal point.
OCP
SET 2.58V 2
.01A
3. Use the wheel knob
to change the output current.
Range
0.00A ~ rating current
Step
10mA
Note
When the output current exceeds the setting
value, PSH automatically switches from CV
(Constant Voltage) to CC (Constant Current)
mode.
PSH User Manual
48
OVP (Output Voltage Protection) Setting
Set the OVP value
1. Press the Menu key
MENU
repeatedly until the OVP menu
appears.
Set OVP
* 21.10V
OCP OUT
CV 2.58V 2.01A
2. Press the F/C key
F/C
to move the
underline before or after the
decimal point.
Set OVP
* 21
.10V
3. Use the wheel knob
to change the value.
4. Press the Iset/Vset/Enter key
VSET/ ISET
ENTER
to confirm the value.
Clear OVP error
1. When the output voltage exceeds
OVP value, the output is shut off
and an OVP error message
appears on the display.
2. Press the Local key
LOCAL
to clear
the error message (the output is
still Off).
OVP Error! Press
“Local” to reset
Range
0.00V ~ rating voltage
Step
10mV (rating voltage < 36V)
20mV (rating voltage 36V)