Calibration
61
Calibration
Run calibration when the output exceeds the
specification, or when the functionality check
(page34) fails. To get the optimal result, make
sure PSH is powered for at least 30 minutes
before the calibration.
Setup
Calibration Preparation ............................. 62
Entering calibration mode ......................... 63
Calibration
Output Voltage calibration......................... 64
Output Current calibration......................... 66
OVP Calibration ....................................... 68
PSH User Manual
62
Calibration Preparation
Calibration
condition
At least 30 minutes of warm-up time
Temperature: 23 ± 5 °C
Relative Humidity: 80%
Calibration
items
Minimum/ Maximum Output Voltage
Minimum/ Maximum Output Current
OVP functionality
Digital
Multimeter
DC Voltage Accuracy: <±0.1%
Recommended model:
GDM-8245, GDM-8246
Current Shunt Current range: >100A
Accuracy: <±0.1%
PSH – Current
Shunt cable
Voltage rating: >70V
Current rating: >100A
Calibration
Equipment
Multimeter –
Current Shunt
cable
N/A