Calibration
63
Entering calibration mode
1. Press the Local key for 5 seconds.
The password entry menu
appears.
Please enter the
Password: 0
2. Enter the model number as the
password.
Wheel knob
changes the digit.
F/C key
F/C
moves the cursor to
the next digit.
(PSH-2018A)
Please enter the
Password: 2018
3. Press the Vset/Iset/Enter key
VSET/ ISET
ENTER
as confirmation.
4. The calibration menu appears.
Wheel knob
selects the item.
Vset/Iset/Enter key
VSET/ ISET
ENTER
confirms the selection.
Minimum/Maximum output
voltage calibration
Minimum/Maximum output
current calibration
Over Voltage Protection calibration
Save the calibration result and exit
the calibration menu
Exit calibration menu without
saving the calibration result
Calibration
Voltage
Calibration
Current
Calibration
OVP
Calibration
Save
Calibration
Exit
PSH User Manual
64
Output Voltage calibration
Connection
Digital Multimeter
PSH Series
V
A
+
Calibration step
1. Enter the Voltage calibration
menu. See page63 for details.
Calibration
Voltage
2. Press the Vset/Iset/Enter
key
VSET/ ISET
ENTER
and enter Minimum
Voltage calibration mode.
Calibration
Voltmin 0.00V
3. Adjust the display value to the
Multimeter reading.
Wheel knob
changes the value.
F/C key
F/C
moves the cursor
before and after the decimal point.
Multimeter reading: 0.02V
Calibration
Voltmin 0.02
V
4. Press the Vset/Iset/Enter
key
VSET/ ISET
ENTER
to confirm the setting.
The display enters Maximum
Voltage calibration mode.
Calibration
Voltmax 20.00V