PSW Series Programming Manual
82
SYSTem:CONFigure:MSLave
Set
Query
Description
Sets or queries the unit operation mode. This
setting is only applied after the unit has been reset.
Syntax
Query Syntax
SYSTem:CONFigure:MSLave { 0 | 1 | 2 | 3 | 4 }
SYSTem:CONFigure:MSLave?
Note
Series mode is only supported for 30V, 80V and
160V models.
Parameter/Return
<NR1>
Description
0
Master/Local
1
Master/Parallel 1 (2 units)
2
Master/Parallel 2 (3 units)
3
Slave/Parallel
4
Slave/Series
SYSTem:CONFigure:OUTPut:EXTernal
[:MODE]
Set
Query
Description
Sets the external logic as active high or active low.
This setting is only applied after the unit has been
reset.
Syntax
Query Syntax
SYSTem:CONFigure:OUTPut:EXTernal[:MODE]
SYSTem:CONFigure:OUTPut:EXTernal[:MODE]?
Parameter
0
Active high
HIGH
Active high
1
Active low
LOW
Active low
Return Parameter
0
<boolean>Active high
1
<boolean>Active low
SYSTem:CONFigure:OUTPut:PON[:STATe]
Set
Query
Description
Sets the unit to turn the output ON/OFF at power-
up. This setting is only applied after the unit has
been reset.