PSW Series User Manual
138
External Operation and Status Monitoring
The MIL 26 pin connector can also be used to
monitor the status operation and alarm status
of the instrument.
The pins are isolated from the power supply
internal circuitry by photo couplers. Status
Com (Pin 17) is a photo coupler emitter output,
whilst pins 18~22 are photo coupler collector
outputs.
A maximum of 30V and 8mA can be applied to
each pin.
Common (photo coupler
emitter) for status signals 18,
19, 20, 21 and 22.
Low when CV mode is active.
Low when CC mode is active.
Low when any of the protection
modes are tripped (OVP,
OCP). Active low.
Low when the output is on.
17
Pins 18, 19, 20, 21, 22