EasyManua.ls Logo

GW Instek PSW 30-72 - Reset to Factory Default Settings

GW Instek PSW 30-72
187 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OPERATION
53
Reset to Factory Default Settings
Background
The F-88 configuration setting allows the PSW
to be reset back to the factory default settings.
See page 161 for the default factory settings.
Steps
1. Press the Function key. The
Function key will light up.
Function
2. The display should show F-
01 on the top and the
configuration setting for
F-01 on the bottom.
3. Rotate the Voltage knob to change
the F setting to F-88 (Factory Set
Value).
Voltage
4. Use the Current knob to set the
F-88 setting to 1 (Return to factory
settings).
Current
5. Press the Voltage knob to confirm.
ConF will be displayed when
successful.
Voltage
6. Press the Function key again to
exit. The function key light will
turn off.
Function
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Other manuals for GW Instek PSW 30-72

Related product manuals