Run Test Script (Automatically at Startup)
The power supply can be configured to
automatically run a test script at startup.
1. Before a test script can be run, it
must first be loaded into one of the
10 memory save slots.
3. Enter the power-on configuration
settings and set F-92 (Power-ON
Output) to run the desired test
script.
4. The selected test script will automatically start
to run the next time the unit is powered on.
*Setting F-92 to 0 or 1 will disable loading a test
script at startup. 0 will turn the output off at
startup. 1 will turn the output on at startup. See
the power on configuration settings for details,
page 113.
When a script is running, press the Output key to
pause the script. To resume the script, press the
Output key again.