CONFIGURATION
109
Bleeder control turns ON/OFF the bleeder
resistor. When set to AUTO the bleeder resistor
is automatically turned on when the output is
turned on and turned off when the output or
power is turned off. See page 27 for usage
details.
When Bleeder Control is turned
OFF or set to AUTO, the
bleeder resistor is turned off
when the power or output is
turned off.
The AUTO setting is only applicable to
firmware version 1.59 or above.
The following table shows how the state of the
bleeder resistor depends on the Bleeder Control
settings, the power state and the output state.
0 = OFF, 1 = ON, 2 = AUTO
Turns the buzzer sound on or off. The buzzer is
associated with alarm sounds and keypad entry
sounds.
Measurement
Average Setting
Determines the level of smoothing for the
average setting.
Only available for firmware version 1.5 or
above.
0 = Low, 1 = Middle, 2 = High