EasyManua.ls Logo

Hach 2100AN - Page 61

Hach 2100AN
108 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SECTION 5, continued
61
Figure 15 illustrates a suggested hookup using a pressure regulator in the inlet
line and a flow meter or flow control valve on the outlet line. The control
valve is installed on the outlet side because it may introduce bubbles into the
sample causing positive interference in the turbidity reading. This hookup also
maintains sufficient pressure to minimize outgassing in a carbonated stream.
Figure 14 Compression Fitting
Figure 15 Suggested High Pressure Installation
Nut
o-ring
Spacer
Grab Ring
Tubing
Sample In
Pressure
Regulator
Flow-Through Cell
Flow Meter
(optional)
Return
Or
To Drain
Control
Valve

Table of Contents

Other manuals for Hach 2100AN

Related product manuals