A fill time setting of 0 seconds causes the instrument to start
measurement immediately.
Use the flow cell specifications in Table 3 to calculate the correct fill time.
Make sure that the fill time includes time to fill the system and to fully
remove the previous sample from the system.
Table 3 Automated flow cell specifications
Specification Details
System flow rate 250 mL/minute
System volume
(from the discharge of the inlet reservoir
to the outlet of the flow cell, not
including the 350 mL inlet reservoir)
30 mL (fill time = 8 seconds)
Purge volume 120 mL (fill time = 30 seconds)
minimum
Recommended to fully remove the
previous sample from the system.
7
Select the measurement time
The measurement time is the time interval that the instrument measures
the sample.
A measurement is completed and the display is updated about once
every second.
Set the measurement time from 0 to 99 minutes and 99 seconds
(minimum=15 seconds).
A measurement time of 0 provides continuous measurement until FLOW
is pushed.
Note: A measurement time of 0 is not recommended for static mode because
particles in the sample may settle over time. The measured turbidity may be lower
than the actual turbidity.
Measurement notes
Before measurement, select the printer to use and the print time interval.
Refer to Configure the printer output on page 41.
7
A shorter fill time may be used when the same sample is being measured again and again.