Table 2 Analyzer data (continued)
Element Definition
REAGENT 2 Level of reagent remaining
REAGENT 3 Level of reagent remaining
STANDARD 1 Level of standard 1 solution remaining
STANDARD 2 Level of standard 2 solution remaining
CLEANER Level of cleaning solution remaining
LEAK COUNTS Indication of possible fluid leaks (range 0 to 1023). A count of over 511 indicates a
fluid leak
Table 3 Measurement data
Element Definition
LAST MEAS CHANNEL Last channel measured.
LAST MEAS TIME Time of last measurement.
LAST ABS
3
Last absorbance reading.
LAST CONC
1
Concentration of last measurement.
NEXT MEAS TIME Time the next measurement will be taken.
DARK
1
Number of A/D counts measured when the LED is turned off.
REF
1
Reference A/D count used to compensate for natural color and turbidity.
SAMPLE
1
A/D counts measurement (after color development) used to determine the
concentration of the sample.
DARK STD DEV
1
Standard deviation of dark counts out of 6 readings.
REF STD DEV
1
Standard deviation of reference counts out of 6 readings.
SAMPLE STD DEV
1
Standard deviation of sample counts out of 6 readings.
REAGENT 1-CELL 1
REAGENT 1-CELL 2
Calculated reagent delivery time to the sample based on temperature, pressure and
viscosity.
REAGENT 2-CELL 1
REAGENT 2-CELL 2
Calculated reagent delivery time to the sample based on temperature, pressure and
viscosity.
REAGENT 3-CELL 1
REAGENT 3-CELL 2
Calculated reagent delivery time to the sample based on temperature, pressure and
viscosity.
Note: Reagent 3 is supplied to both colorimetric cells only during calibration.
View instrument information
1. Select menu and select SETUP SYSTEM>INSTRUMENT INFORMATION.
2. Select an option.
Option Description
ANALYZER INFO Shows the software information and serial number.
MODULE INFO (option is available when a
module is installed)
Shows the connected modules with software information
and serial number.
3
Information is available for both sample cells