WGM - 28 November 2007 - Edition 01
PAT700 Total Organic Carbon Analyzer – Bottle Mode Page 93
Anatel Operator Manual
6.9 Grab sample
CAUTION
Potential puncture hazard! Contact with the exposed needle could cause serious
injury. Use extreme care when installing and removing the sample bottles.
The grab sample function enables analysis of samples from various locations. Up to three
repetitions from a single sample may be analyzed to arrive at an average value. Grab samples
are run using a standard water sample bottle and do not require the use of an RFID tagged
bottle.
1) At the home screen, touch the << icon to open the sliding toolbar, then touch the
bottles icon in the toolbar, or touch the bottles icon in the lower left corner of the
screen.
2) Touch the grab sample button in the bottle mode dialog box. See Figure 6-1,
page 76.
3) The OASIS door will unlock for five seconds. If the door is not opened within five
seconds, the unlock icon can be used to unlock the door manually.
4) Install the grab sample bottle following the instructions as shown in the load bottles
dialog box and ensuring that you place the grab sample bottle in the correct
position. See Figure 6-19. From left:
• Bottle 1 = empty
• Bottle 2 = sample bottle
• Bottle 3 = empty
• Bottle 4 = empty
Figure 6-19 Load bottles dialog box, grab sample
• Bottle positions graphics show you where to install the bottle.
• Unlock icon energizes the bottle bay door lock for five seconds, allowing you to
open the door.