WGM - 28 November 2007 - Edition 01
PAT700 Total Organic Carbon Analyzer – Bottle Mode Page 95
Anatel Operator Manual
Figure 6-21 Grab sample test in progress dialog box
8) When the grab sample test in progress dialog box appears, the analyzer begins
running the test.
Upon starting the grab sample, the analyzer performs a temperature stabilization.
Once the temperature has stabilized, the sample bottle is analyzed.
Cancel icon takes you to the unload/replace bottles dialog box. All changes made
in any of the dialog boxes are lost.
9) When the test is completed, the grab sample summary dialog box appears, with the
results of the test. The results also appear on the printout if a printer is attached.
See Figure 6-22.
Figure 6-22 Grab sample summary dialog box
Export icon opens the export grab sample dialog box. See Figure 6-23. From there, you can
export the grab sample summary to a USB flash drive
Run another bottle icon returns you to the load bottles dialog box.
Cancel icon takes you to the unload/replace bottles dialog box. See Figure 6-11, page 84.