EasyManua.ls Logo

Hach ANATEL PAT700 - Excursion with Validation

Hach ANATEL PAT700
158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WGM - 28 November 2007 - Edition 01
PAT700 Total Organic Carbon Analyzer – Bottle Mode Page 97
Anatel Operator Manual
Figure 6-24 Excursion mode dialog box
Excursion with validation button opens the excursion with validation load bottles
dialog box. See Figure 6-25 and follow the instructions in Excursion with validation,
below.
Excursion without validation button tells the analyzer to run the excursion without
validation and opens the excursion without validation load bottles dialog box. See
Figure 6-29 and follow the instructions in Excursion without validation, page 100.
Return icon returns you to the bottle mode dialog box.
Excursion with validation
CAUTION
Potential puncture hazard! Contact with the exposed needle could cause serious
injury. Use extreme care when installing and removing the sample bottles.
1) Install the standards bottles following the instructions as shown in the load bottles
dialog box and ensuring that you place the standards bottle in the correct position.
See Figure 6-25. From left:
Bottle 1 = blank
Bottle 2 = empty
Bottle 3 = 500 ppb sucrose
Bottle 4 = excursion bottle

Table of Contents

Other manuals for Hach ANATEL PAT700

Related product manuals