Page 106 Alarms – PAT700 Total Organic Carbon Analyzer
Operator Manual Anatel
WGM - 28 November 2007 - Edition 01
Error alarms
Errors occur any time one of the non-critical subsystems fail or a failure occurs from which the
analyzer attempts to recover. The animation area flashes red to indicate an error condition is
present. When an error condition is present, the analyzer attempts to recover and continue
operating. See Table 7-1.
Table 7-1 Error alarm messages
Code Message Alarm condition
108 UnableToReadSettings Some or all of the settings could not be read from the settings file.
109 UnableToWritetoSettings Analyzer cannot write to settings file.
112 UnableToStartPrinterModule Analyzer cannot start the printer module.
113 UnableToStartRS232Module Analyzer cannot start the RS-232 module.
115 UnableToStartLamp Analyzer cannot start the lamp module.
116 UnableToReadDataLog Analyzer cannot read data log.
117 UnableToWriteToDataLog Analyzer cannot write to data log.
123 UnableToReadAlarmLog Analyzer cannot read alarm log.
124 UnableToWritetoAlarmLog Analyzer cannot write to alarm log.
128 UvLampMonitorFailed UV lamp monitor failed.
140 UnableToWriteAuditTrailEntry Analyzer cannot write audit trail entry.
161 TocAlarm TOC value has exceeded the configured TOC alarm limit.
162 ConductivityAlarm Conductivity reading exceeds the USP <645> conductivity limits.
163 UnableToStopCurrentOperation Analyzer cannot stop its current operation.
164 UnknownError Analyzer has experienced an unknown error.
167 WatchDogIOCard I/O Card has stopped responding.
168 WatchDogTocAlgoithm TOC algorithm has stopped responding.
169 WatchDogPrinter Printer has stopped responding.
170 WatchDogRs232 RS-232 module has stopped responding.
171 WatchDogModbus Modbus module has stopped responding.
177 Exception An exception was thrown.