PAT700 Total Organic Carbon Analyzer – Alarms Page 107
Anatel Operator Manual
Critical error alarms
A critical error occurs any time the analyzer cannot function due to an error in the system. The
error appears as a full-screen, flashing message. All operation ceases until the error condition
has been corrected. The current analysis is aborted. Touching the screen provides access to
the alarm acknowledge dialog box. See Table 7-2.
Warning alarms
A warning occurs any time one of the subsystems is not operating at a normal level. When a
warning condition is present, a warning indicator appears on the display. Operation may
continue with a warning condition present. When the warning condition has been corrected,
the indicator disappears. See Table 7-3.
Table 7-2 Critical error alarm messages
Code Message Alarm condition
110 UnableToStartIoModule Analyzer cannot start IO module.
111 UnableToStartTocModule Analyzer cannot start TOC module
114 UnableToStartEthernetModule Analyzer cannot start Ethernet module.
152 LampUnableToSwitch Analyzer cannot switch from one lamp to the other.
166 UnableToInitializePump Failed to initialize pump to correct speed.
173 RangeTable_FailedToLoad RangeTable.txt file could not be loaded.
174 AnalyzerTemperature Temperature inside analyzer is out of normal operating range.
175 BothLampsBad Both UV lamps have failed.
Replace both lamps as soon as possible.
176 NoCondTempReadings Analyzer cannot measure conductivity and temperature.
Restart analyzer.
If restarting doesn't resolve error, phone customer service at
1–800–866–8854.
179 HardwareInitializationFailure An error occurred during hardware initialization.
Table 7-3 Warning alarm messages
Code Message Alarm condition
27 FALSE_CONDUCTIVITY Measured conductivity is less than 0.
28 BAD_TEMP_CONVERT Conductivity out of bounds for temperature conversion.
29 BAD_TOC_CONVERT
34 ANALYSIS_TIMEOUT Unable to finish oxidation in 35 minutes.
35 LOW_OX_RATE TOC too low (out of range).
36 CHEM_INTERFERE Negative TOC (out of range).
Potential presence of organic acids.