Page 108 Alarms – PAT700 Total Organic Carbon Analyzer
Operator Manual Anatel
WGM - 28 November 2007 - Edition 01
37 TEMP_LOW Temperature out of range low.
38 TEMP_HIGH Temperature out of range high.
39 RESIST_LOW Resistivity out of range low.
40 RESIST_HIGH Resistivity out of range high.
100 IncorrectBottleLoadedinPosition1 • Incorrect bottle detected in position 1.
• Analyzer cannot perform bottle test operation.
101 IncorrectBottleLoadedinPosition2 • Incorrect bottle detected in position 2.
• Analyzer cannot perform bottle test operation.
102 IncorrectBottleLoadedinPosition3 • Incorrect bottle detected in position 3.
• Analyzer cannot perform bottle test operation.
103 IncorrectBottleLoadedinPosition4 • Incorrect bottle detected in position 4.
• Analyzer cannot perform bottle test operation.
104 NotEnoughFluidinBottle1 • Not enough fluid in bottle 1 to complete operation.
• Check bottle 1 before starting operation.
• Analyzer aborts operation and returns to its normal operating mode.
105 NotEnoughFluidinBottle2 • Not enough fluid in bottle 2 to complete operation.
• Check bottle 2 before starting operation.
• Analyzer aborts operation and returns to its normal operating mode.
106 NotEnoughFluidinBottle3 • Not enough fluid in bottle 3 to complete operation.
• Check bottle 3 before starting operation.
• Analyzer aborts operation and returns to its normal operating mode.
107 NotEnoughFluidinBottle4 • Not enough fluid in bottle 4 to complete operation.
• Check bottle 4 before starting operation.
• Analyzer aborts operation and returns to its normal operating mode.
118 UnableToChangeLanguageSettings Analyzer cannot change language settings.
119 UnableToReadLanguageSettings Analyzer cannot read language settings
120 UnableToChangeEthernetSettings Analyzer cannot change Ethernet settings.
121 UnableToReadEthernetSettings Analyzer cannot read Ethernet settings.
122 UnableToPrintSystemSettings Analyzer cannot print system settings.
125 UnableToStartLampTest Analyzer cannot start lamp test.
126 UnableToStopLampTest Analyzer cannot stop lamp test.
127 UvLampLifeExceeded • Lamp hours of usage have been exceeded or UV Detect™ output
has dropped below acceptable limit.
• Lamp has been set to a status of marginal.
• Replace lamp.
129 UnableToReadInformationFromBottle Analyzer cannot read information from bottle.
130 UnableToWriteInformationToBottle Analyzer cannot write information to bottle.
Table 7-3 Warning alarm messages
Code Message Alarm condition